User manual
Section
7: TSP command reference Series 2600B System SourceMeterĀ® Instrument
Reference Manual
7-368 2600BS-901-01 Rev. B / May 2013
Example 1
instrumentID = tspnet.connect("192.0.2.1")
if instrumentID then
-- Use instrumentID as needed here
tspnet.disconnect(instrumentID)
end
Connect to a TSP-
enabled
device.
Example 2
instrumentID = tspnet.connect("192.0.2.1", 1394,
"*rst\r\n")
if instrumentID then
-- Use instrumentID as needed here
tspnet.disconnect(instrumentID)
end
Connect to a device that is
not TSP-enabled.
Also see
localnode.prompts (on page 7-141)
localnode.showerrors (on page 7-145)
tspnet.tsp.abortonconnect (on page 7-374)
tspnet.disconnect() (on page 7-368)
tspnet.disconnect()
This function disconnects a specified TSP-Net session.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
tspnet.disconnect(connectionID)
connectionID
The connection ID returned from tspnet.connect()
Details
This function disconnects the two devices by closing the connection. The connectionID is the session handle
returned by tspnet.connect().
For TSP-enabled devices, this aborts any remotely running commands or scripts.
Example
testID = tspnet.connect("192.0.2.0")
-- Use the connection
tspnet.disconnect(testID)
Create a TSP-Net session.
Close the session.
Also see
tspnet.connect() (on page 7-367)