User manual
Section
7: TSP command reference Series 2600B System SourceMeterĀ® Instrument
Reference Manual
7-344 2600BS-901-01 Rev. B / May 2013
trigger.timer[N].clear()
This function clears the timer event detector and overrun indicator for the specified trigger timer number.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.timer[N].clear()
N
Trigger timer number to clear (1 to 8)
Details
This command sets the timer event detector to the undetected state and resets the overrun indicator.
Example
trigger.timer[1].clear()
Clears trigger timer 1.
Also see
trigger.timer[N].count (on page 7-344)
trigger.timer[N].count
This attribute sets the number of events to generate each time the timer generates a trigger event.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Trigger timer N reset
Not saved
1
Usage
count = trigger.timer[N].count
trigger.timer[N].count = count
count
Number of times to repeat the trigger (0 to 1,048,575)
N
A trigger timer number (1 to 8)
Details
If count is set to a number greater than 1, the timer automatically starts the next delay at expiration of the
previous delay.
Set count to zero (0) to cause the timer to generate trigger events indefinitely.
Example
print(trigger.timer[1].count)
Read trigger count for timer number 1.
Also see
trigger.timer[N].clear() (on page 7-344)
trigger.timer[N].reset() (on page 7-348)