User manual
Series 2600B
System SourceMeterĀ® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-245
Usage
eventID = smuX.trigger.MEASURE_COMPLETE_EVENT_ID
eventID
The measurement complete event number
X
Source-measure unit (SMU) channel (for example,
smua.trigger.MEASURE_COMPLETE_EVENT_ID
applies to SMU channel A)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to measure
complete events from this SMU.
Also see
Triggering (on page 3-32)
smuX.trigger.PULSE_COMPLETE_EVENT_ID
This constant contains the pulse complete event number.
Type TSP-Link accessible Affected by Where saved Default value
Constant
Yes
Usage
eventID = smuX.trigger.PULSE_COMPLETE_EVENT_ID
eventID
The pulse complete event number
X
Source-measure unit (SMU) channel (for example,
smua.trigger.PULSE_COMPLETE_EVENT_ID
applies to SMU channel A)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to pulse
complete events from this SMU.
Also see
Triggering (on page 3-32)