User manual
Series 2600B
System SourceMeterĀ® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-137
Example
lan.trigger[5].stimulus = trigger.timer[1].EVENT_ID
Use timer 1 trigger event as
the source for LAN packet 5
trigger stimulus.
Also see
lan.trigger[N].assert() (on page 7-128)
lan.trigger[N].clear() (on page 7-129)
lan.trigger[N].connect() (on page 7-130)
lan.trigger[N].overrun (on page 7-134)
lan.trigger[N].wait() (on page 7-137)
lan.trigger[N].wait()
This function waits for an input trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
triggered = lan.trigger[N].wait(timeout)
triggered
Trigger detection indication
N
The trigger packet over LAN to wait for (1 to 8)
timeout
Maximum amount of time in seconds to wait for the trigger event
Details
If one or more trigger events have been detected since the last time lan.trigger[N].wait() or
lan.trigger[N].clear() was called, this function returns immediately.
After waiting for a LAN trigger event with this function, the event detector is automatically reset and rearmed
regardless of the number of events detected.
Example
triggered = lan.trigger[5].wait(3)
Wait for a trigger with LAN packet 5 with a timeout of
3 seconds.
Also see
lan.trigger[N].assert() (on page 7-128)
lan.trigger[N].clear() (on page 7-129)
lan.trigger[N].overrun (on page 7-134)
lan.trigger[N].stimulus (on page 7-135)
localnode.autolinefreq
This attribute enables or disables automatic power line frequency detection at start-up.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Not applicable
Nonvolatile memory
true (enabled)