display.smuX.digits
This attribute sets the display resolution of the selected measurement.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Instrument reset
Recall setup
Saved setup
5 (display.DIGITS_5_5)
Usage
digits = display.smuX.digits
display.smuX.digits = digits
digits
Set digits to one of the following values:
Select 4-1/2 digit resolution (4 or display.DIGITS_4_5)
Select 5-1/2 digit resolution (5 or display.DIGITS_5_5)
Select 6-1/2 digit resolution (
6
or
display.DIGITS_6_5)
X
Source-measure unit (SMU) channel (for example, display.smua.digits
applies to SMU channel A)
Details
SMU A and SMU B can be set for different measurement display resolutions.
Example
display.smua.digits = display.DIGITS_5_5
Select 5-1/2 digit resolution for SMU A.
Also see
Display resolution (on page 3-72)