3: Functions and features Series 2600B System SourceMeterĀ® Instrument
Reference Manual
3-14 2600BS-901-01 Rev. B / May 2013
Buffer read-only attributes
The following table contains buffer read-only attributes.
Buffer read-only attributes: Read-only attributes used to access buffer parameters
Storage attribute Description
basetimestamp
midnight January 1, 1970 GMT. See Time and date values (on page 7-3) for
capacity
n
next
Buffer storage control programming examples
The programming examples below illustrate the use of buffer storage control attributes.
Buffer control programming examples:
Command Description
smua.nvbuffer1.collectsourcevalues = 1
Enable source value storage.
smua.nvbuffer1.appendmode = 1
Enable buffer append mode.
smua.nvbuffer1.collecttimestamps = 0
Disable timestamp storage.
smua.nvbuffer1.timestampresolution =
0.001
Set timestamp resolution to 0.001024 s.
smua.nvbuffer1.fillcount = 50
Set 50 as the number of readings the
buffer will take before restarting at index
1.
smua.nvbuffer1.fillmode = 0
Set the reading buffer to fill once (do
not overwrite old data).
Buffer read-only attribute programming examples
The follow programming examples illustrate use of buffer read-only attributes.
Buffer read-only attribute programming examples:
Command Description
number = smua.nvbuffer1.n
Request the number of readings in the
buffer.
buffer_size = smua.nvbuffer1.capacity
Request buffer size.