Data Sheet
TEK.COM 11
2600B System SourceMeter
®
 SMU Instruments
Contact Check
 1
 (not available on Model 2604B)
Speed
Maximum Measurement Time To Memory 
For 60Hz (50Hz)
Accuracy (1 Year), 23 °C ±5 °C 
±(%rdg. + ohms)
FAST 1 (1.2) ms 5% + 10 W
MEDIUM 4 (5) ms 5% + 1 W
SLOW 36 (42) ms 5% + 0.3 W
NOTES
1.  Includes measurement of SENSE HI to HI and SENSE LO to LO contact resistances.
Additional Meter Specifications
Maximum Load Impedance  Normal Mode: 10 nF (typical). High Capacitance Mode: 50 µF (typical).
Common Mode Voltage  250 VDC.
Common Mode Isolation  >1 G W, <4500 pF.
Overrange  101% of source range, 102% of measure range.
Maximum Sense Lead Resistance  1 kW for rated accuracy.
Sense Input Impedance  >10 G W.
High Capacitance Mode
 1, 2, 3
Voltage Source Output  Time required to reach 0.1% of final value after source level command is processed on a fixed range.
Settling Time  Current limit = 1 A.
Voltage Source Range Settling Time with C
load
 = 4.7 µF
100 mV 200 µs (typical)
1 V 200 µs (typical)
6 V 200 µs (typical)
40 V  7 ms (typical)
Current Measure Settling Time  Time required to reach 0.1% of final value after voltage source is stabilized on a fixed range. Values below 
for V
out
 = 1V unless noted.
Current Measure Range Settling Time
3 A – 1 A <120 µs (typical) (R
load
 > 2 W)
100 mA – 10 mA <100 µs (typical)
1 mA < 3 ms (typical)
100 µA < 3 ms (typical)
10 µA < 230 ms (typical)
1 µA < 230 ms (typical)
Capacitor Leakage Performance Using HIGH-C Scripts
 4
Load = 5 µF||10 MW. Test: 5 V step and measure. 200 ms (typical) @ 50 nA.
NOTES
1.  High Capacitance Mode specifications are for DC measurements only. 
2.  100 nA range is not available in High Capacitance Mode.
3.  High Capacitance Mode utilizes locked ranges. Auto Range is disabled.
4.  Part of KI Factory scripts. See reference manual for details.