Datasheet
Datasheet
TEK.COM8
Bias Current Source Specifications
Current Programming Accuracy
Range
Programming
Resolution
Accuracy (1 Year)
±(% rdg. + amps)
Typical Noise (RMS)
0.1 Hz–100 kHz
250 mA 10 µA 0.17% + 1 mA 200 µA
Temperature Coefficient (0°–18 °C and 28°–50 °C)
± (0.15 × accuracy specification)/°C.
Pulse Width
Programming Resolution 1 µs.
Pulse Width
Programming Maximum 500 µs.
Pulse Width
Programming Minimum 10 µs.
Pulse Width
Programming Accuracy ±200 ns.
Pulse Width Jitter 110 ns (typical).
Pulse Period Jitter 2 µs (typical).
Pulse Width Programming,
Minimum t
off
16 μs.
Pulser Feature Measure Specifications
Voltage Measurement Specifications
Voltage Measurement Accuracy
Range Display Resolution
Accuracy (1 Year)
23 °C ± 5 °C
± (% rdg. + volts)
1
5 V 1 µV 0.05% + 2.5 mV
10 V 10 µV 0.05% + 4 mV
Voltage Measurement Time required to reach specified accuracy after a source level command is processed on a fixed range.
Settling Time
Voltage Range Settling Time (typical)
5 V and 10 V < 9 µs
Temperature Coefficient (0°–18 °C and 28°–50 °C)
± (0.15 × accuracy specification)/C.