Datasheet

TEK.COM 7
2601B-PULSE System SourceMeter 10 µs Pulser/SMU Instrument
Additional Pulser Source Specifications
Rise Time (10% to 90%) <1.7 µs for a full-scale step of current into any load voltage (10V maximum).
200 (ns × A)
Additional Zero-Crossing Delay 1 μs +
_______________________
Pulse Current (in Amps)
Pulse Current and Duty Cycle Maximum duty cycle is given by:
0.3125 – |I
bias
|
*
100%
||I
pulse
| – |I
bias
||
For
I
bias
≤ 10 mA:
±10 A
±5 A
±3 A
±1 A
±500 mA
±250 mA
0 A
3 %
Pulse Current
Max. Duty Cycle
6%
10 %
30 %
60 %
100 %
Current Source Output Settling Time
Time required to reach specified accuracy after the start of the pulse.
Current Range Settling Time
1 A < 9 µs (V
load
≤ 10 V)
5 A < 9 µs (V
load
≤ 10 V)
10 A < 9 µs (V
load
≤ 10 V)
Output Off Normal State Electrical short (< 1 Ω) between HI and LO.
Maximum DC current from external sources during OUTPUT OFF state must be limited to <1 A.
Remote Voltage Sense Maximum voltage between HI and SENSE HI = ±30 V.
Maximum voltage between LO and SENSE LO = ±30 V.
Over Temperature Protection Internally sensed temperature overload puts unit in standby mode.
Safety Interlock Hardware interlock (available, optional).