Datasheet
TEK.COM 17
2601B-PULSE System SourceMeter 10 µs Pulser/SMU Instrument
USB File System (Front) USB 2.0 Host: Mass storage class device.
Power Supply 100 V to 240 VAC, 50–60 Hz (auto sensing), 240 VA maximum.
Cooling Forced air. Side intake and rear exhaust. One side must be unobstructed when rackmounted.
EMC Conforms to European Union EMC Directive.
Safety NRTL listed to UL61010-1:2008 and CSA C22.2 No. 61010-1. Conforms to European Union Low Voltage
Directive.
Dimensions
2601B-PULSE only 105 mm × 235 mm × 445 mm (4.1 in. high × 9.25 in. wide × 17.5 in. deep).
2601B-PULSE with
2601B-P-INT attached 105 mm × 235 mm × 503 mm (4.1 in. high × 9.25 in. wide × 19.82 in. deep).
Weight 2601B-PULSE only: 5.9 kg (13 lb).
2601B-PULSE with 2601B-P-INT attached: 6.4 kg (14 lb).
Environment For indoor use only.
Altitude Maximum 2000 meters above sea level.
Operating 0 °C to 35 °C at up to 70% relative humidity; at 35 °C to 50 °C, derate 3% relative humidity per °C.
Storage –25 °C to 65 °C.
Supplemental Information
Front Panel Interface Two-line vacuum fluorescent display (VFD) with keypad and rotary knob.
Display Show error messages and user defined messages.
Display source and limit settings.
Show current and voltage measurements.
View measurements stored in dedicated reading buffers.
Keypad Operations Change host interface settings.
Save and restore instrument setups.
Load and run factory and user defined test scripts (i.e. sequences) that prompt for input and send results
to the display.
Store measurements into dedicated reading buffers.
Programming Embedded Test Script Processor (TSP
®
) accessible from any host interface; responds to high-speed test
scripts comprised of remote commands and statements (for example, branching, looping, and math); able
to execute test scripts stored in memory without host intervention.
Minimum Memory Available 16 MB (approximately 250,000 lines of TSP code).
Reading Buffers Nonvolatile memory uses dedicated storage areas reserved for measurement data. Reading buffers are
arrays of measurement elements. Each element can store the following items:
Measurement
Source setting (at the time the measurement was made)
Measurement status
Range information
Timestamp
Reading buffers can be filled using the front panel STORE key and retrieved using the RECALL key or
hostinterface.
Buffer Size, with timestamp and source setting: >60,000 samples.
Buffer Size, without timestamp and source setting: >140,000 samples.
Timer Free running 47-bit counter with 1 MHz clock input. Reset each time instrument powers up. If the
instrument is not turned off, the timer is automatically reset to zero (0) every four years.
Timestamp TIMER value automatically saved when each measurement is triggered.
Resolution 1 µs.
Accuracy ±100 ppm.