Datasheet

TEK.COM 13
2601B-PULSE System SourceMeter 10 µs Pulser/SMU Instrument
Additional Characteristics
Maximum Load Impedance
Normal Mode 10 nF.
High Capacitance Mode 50 µ F.
Overrange 101% of source range, 102% of measure range.
Maximum Sense Lead Resistance 1 kW for rated accuracy.
Sense High Input Impedance >10 G W.
High Capacitance Mode
1, 2, 3
Accuracy Specifications Accuracy specifications are applicable in both Normal and High Capacitance Modes.
Voltage Source Output Time required to reach 0.1% of final value after source level command is processed on a fixed range.
Settling Time Current limit = 1 A.
Voltage Source Range Settling Time with C
load
= 4.7 μF
100 mV <200 μs
1 V <200 μs
6 V <200 μs
40 V <7 ms
Current Measure Settling Time Time required to reach 0.1% of final value after voltage source is stabilized on a fixed range. Values below
for V
out
= 1V unless noted.
Current Range Settling Time
1 μA < 230 ms
10 μA < 230 ms
100 μA < 3 ms
1 mA < 3 ms
10 mA – 100 mA <100 μs
1 A – 3 A
<120 μs (R
load
> 2 W)
Capacitor Leakage Performance Using the KIHighC Factory Script
4
Load = 5 μF||10 MW. Test: 5 V step and measure. 200 ms (typical) @ 50 nA.
NOTES
1. High Capacitance Mode specifications are for DC measurements only.
2. 100 nA range is not available in High Capacitance Mode.
3. High Capacitance Mode uses locked ranges. Auto Range is disabled.
4. Part of KI Factory scripts. See reference manual for details.
Mode Change Delay
Current Ranges of Delay into High Capacitance Mode: 11 ms.
100 μA and Above Delay out of High Capacitance Mode: 11 ms.
Current Ranges
Delay into High Capacitance Mode: 250 ms.
Below 100 μA
Delay out of High Capacitance Mode: 11 ms.
Voltmeter Input Impedance 10 GW in parallel with 3300 pF.
Noise, 10 Hz–20 MHz (6 V Range) <30 mV peak-peak (typical).
Voltage Source Range Change Overshoot
<400 mV + 0.1% of larger range. Overshoot into a 100 kW load, 20 MHz bandwidth.