Datasheet

TEK.COM 11
2601B-PULSE System SourceMeter 10 µs Pulser/SMU Instrument
Additional Source Characteristics
Noise, 10 Hz to 20 MHz <20 mV peak-peak, < 3 mVRMS.
6 V range.
Transient Response Time < 70 μs for the output to recover to within 0.1% for a 10% to 90% step change in load.
Overshoot
Voltage <±0.1% + 10 mV. Step size = 10% to 90% of range, resistive load, maximum current limit/compliance.
Current <±0.1%. Step size = 10% to 90% of range, resistive load.
See Current Source output settling time for additional test conditions.
Range Change Overshoot
Voltage
1
<300 mV + 0.1% of larger range. Overshoot into a 100 kΩ load, 20 MHz bandwidth
Current
2
<300 mV/R
LOAD
+ 5% of larger range.
Guard Offset Voltage < 4 mV. Current <10 mA.
Remote Sense Operating Range
3
Maximum voltage between HI and SENSE HI = 3 V.
Maximum voltage between LO and SENSE LO = 3 V.
Voltage Output Headroom
40 V Range Maximum output voltage = 42 V – (total voltage drop across source leads). Maximum 1 Ω per source lead.
6 V Range Maximum output voltage = 8 V – (total voltage drop across source leads). Maximum 1 Ω per source lead.
Overtemperature Protection Internally sensed overtemperature condition puts the instrument in standby mode.
Limit/Compliance Bipolar limit (compliance) set with a single value.
Voltage
4
Minimum value is 10 mV; accuracy is the same as voltage source.
Current
5
Minimum value is 10 nA; accuracy is the same as current source.
Voltage Source Output Settling Time
Time required to reach within 0.1% of final value after source level command is processed on a fixed range.
Voltage Range Settling Time
100 mV < 50 μs
1 V < 50 μs
6 V < 110 μs
40 V
6
< 150 μs
Current Source Output Settling Time
Time required to reach within 0.1% of final value after source level command is processed on a fixed range.
Values below for I
OUT
× R
LOAD
= 1 V unless noted.
Current Range Settling Time
100 nA < 20 ms
1 μA < 2 ms
10 μA < 500 μs
100 μA < 150 μs
1 mA < 100 μs
10 mA to 1 A < 80 μs (R
LOAD
> 6 Ω)
3 A < 80 μs (Current < 2.5 A, R
LOAD
> 2 Ω)
NOTES
1. Add 180 mV for the 6 V to 20 V change (500 mV).
2. With source settling set to SETTLE_SMOOTH_100NA.
3. Add 50 μV to source accuracy specifications per volt of HI lead drop.
4. For sink operation (quadrants II and IV) without sink mode enabled, add 10% of compliance range and ±0.02% of limit settling to the corresponding voltage source accuracy specifications.
For the 100 mV range, add an additional 60 mV of uncertainty. Specifications apply with sink mode enabled.
5. For sink operation (quadrants II and IV) without sink mode enabled, add 0.06% of limit range to the corresponding current limit accuracy specifications. Specifications apply with sink mode
enabled.
6. Add 150 μs when measuring on the 1 A range.