Datasheet

Datasheet
TEK.COM10
SMU Specifications
Voltage Source Specifications
Voltage Accuracy
1
Range
Programming
Resolution
Accuracy (1 Year)
±(% rdg. + volts)
Typical Noise (peak-peak)
0.1 Hz–10 Hz
100 mV 5 µV 0.02% + 250 µV 20 µV
1 V 50 µV 0.02% + 400 µV 50 µV
6 V 50 µV 0.02% + 1.8 mV 100 µV
40 V 500 µV 0.02% + 12 mV 500 µV
Temperature Coefficient (0°–18 °C and 28°–50 °C)
2
±(0.15 × accuracy specification)/ °C. Applicable for normal mode only. Not applicable for High
CapacitanceMode.
Maximum Output Power and Source/Sink Limits
3
40.4 W maximum. ±40.4 V @ ±1.0 A, ±6.06 V @ ±3.0 A, four quadrant source or sinkoperation.
NOTES
1. Add 50 µV to source accuracy specifications per volt of HI lead drop.
2. High Capacitance Mode accuracy is applicable at 23 °C ±5 °C only.
3. Full power source operation regardless of load to 28 °C ambient. Above 28 °C and/or power sink operation, refer to “Operating Boundaries” in the 2601B-PULSE Reference Manual for
additional power derating information.
Current Source Specifications
Current Accuracy
Range
Programming
Resolution
Accuracy (1 Year)
±(% rdg. + amps)
Typical Noise (peak-peak)
0.1Hz–10Hz
100 nA 2 pA 0.1% + 100 pA 5 pA
1 µA 20 pA 0.03% + 800 pA 25 pA
10 µA 200 pA 0.03% + 5 nA 60 pA
100 µA 2 nA 0.03% + 60 nA 3 nA
1 mA 20 nA 0.03% + 300 nA 6 nA
10 mA 200 nA 0.03% + 6 µA 200 nA
100 mA 2 µA 0.03% + 30 µA 600 nA
1 A 20 µA 0.05% + 1.8 mA 70 µA
3 A 20 µA 0.06% + 4 mA 150 µA
10 A
2
200 µA 0.5% + 40 mA N/A
Temperature Coefficient (0°–18 °C and 28°–50 °C)
3
±(0.15 × accuracy specification)/ °C.
Maximum Output Power and Source/Sink Limits
1
40.4 W maximum. ±1.01 A @ ±40.0 V, ±3.03 A @ ±6.0 V, four quadrant source or sinkoperation.
NOTES
1. Full power source operation regardless of load to 28 °C ambient. Above 28 °C and/or power sink operation, refer to “Operating Boundaries” in the 2601B-PULSE Reference Manual for
additional power derating information.
2. 10 A Range is accessible in SMU extended range mode only. Accuracy specifications for 10 A range are typical.
3. High Capacitance Mode accuracy is applicable at 23 °C ±5 °C only.