2400 Series SourceMeter
®
 User’s Manual Connections 2-7
sensing must be selected. The 2-wire sense mode is the BENCH and 
GPIB default. See 
“Sense selection,” page 2-12, for details.
4-wire remote sensing
When sourcing and/or measuring voltage, there can be errors associated with IR 
drops in the test leads. Voltage source and measure accuracy are optimized by 
using 4-wire remote sense connections. When sourcing voltage, 4-wire remote 
sensing ensures that the programmed voltage is delivered to the DUT. When 
measuring voltage, only the voltage drop across the DUT is measured. Use 4-wire 
remote sensing for the following source-measure conditions:
• Test circuit impedance is <1kΩ.
• Optimum Ohms, V-Source, and/or V-Measure accuracy are required.
NOTE Specified accuracies for both source and measure are only achieved using 
4-wire remote sensing.
WARNING When sourcing voltage in remote sense, make sure the sense 
leads are connected to the DUT. If a sense lead becomes dis
-
connected, 0V will be sensed, and the SourceMeter will 
increase the output voltage (to possibly hazardous levels) to 
compensate. For increased safety, you can limit the voltage 
output by the SourceMeter. See 
Section 3, “V-source protec-
tion,” to limit the output voltage level.
NOTE When the output is turned off in the 4-wire sense mode, the sense lines 
are internally disconnected for safety reasons. The sense lines are auto
-
matically reconnected when the output is turned on, and 4-wire sensing is 
restored.
2-wire local sensing
2-wire local sense connections can only be used if the error contributed by test 
lead IR drop is acceptable to the user. At current levels below 100mA, the errors 
are usually not significant (assuming test lead resistance is not greater than 1Ω).
Since current in a series circuit is the same at all points in the loop, remote sens-
ing does not improve I-Source or I-Measure accuracy. Thus, if sourcing current 
and measuring current, you can use local sensing. Other conditions that allow the 
use of 2-wire local sensing include:
• Test circuit impedance is 1kΩ. Above 1GΩ, guarding should also be used  
(“Cable guard”).
• Measure Only (V or I) operation.