2400 Series SourceMeter
®
 User’s Manual Specifications A-3
Source-Delay-Measure (SDM) cycle timing
The following timing information assumes that the SourceMeter is being triggered 
externally via the Trigger Link. 
For Cases I through IV, it is assumed that the Output Auto-Off feature is enabled 
(:SOURce1:CLEar:AUTO ON), and the source setting changes for each triggered 
SDM cycle. The discussion is applicable for linear, log, and custom sweeps. It is 
also applicable to applications that use the “triggered source” feature 
(:SOURce1:VOLTage:TRIGger or SOURce1:CURRent:TRIGger). The discussion 
is not applicable for memory sweeps (:SOURce1:MEMory). 
For Cases V and VI, it is assumed that the Output Auto-Off feature is disabled 
(:SOURce1:CLEar:AUTO OFF), and the source setting remains the same for 
each triggered SDM cycle. In this configuration, the static source remains on dur-
ing all SDM cycles.
Definitions
Trigger latency
Trigger latency is the time from when an external trigger event occurs to when the 
Source-Meter takes the appropriate action. It is from when an external trigger is 
detected in the Trigger Layer of the trigger model to when the trigger delay begins. 
Trigger delay
Trigger delay is the time from when the external event is detected to when the 
source configuration begins. This is a user-programmable delay that can be set 
from 0000.0000 seconds to 999.99990 seconds.
Source configuration
This is the time it takes to configure the source DAC. For the following discussion, 
it is assumed that the range and polarity do not change when the source value is 
changed. 
Source delay
This is the time between the source configuration and the start of the first A/D con-
version. This programmable delay is typically used to allow the source to settle 
before starting the measurement. With Auto-Delay enabled, 100µsec is added to 
the user-programmed source delay. The user-programmed delay can be set from 
0000.0000 seconds to 9999.99900  