User Manual
2400 Series SourceMeter
®
 User’s Manual SCPI Command Reference 18-101
STATus subsystem
The STATus subsystem is used to control the status registers of the SourceMeter. 
The commands in this subsystem are summarized in Table 18-8.
NOTE These registers and the overall status structure are fully explained in 
Section 15, “Status Structure.”
Read event registers
[:EVENt]?
:STATus:MEASurement[:EVENt]? Read Measurement Event Register
:STATus:QUEStionable[:EVENt]? Read Questionable Event Register
:STATus:OPERation[:EVENt]? Read Operation Event Register
Description These query commands are used to read the contents of the status 
event registers. After sending one of these commands and address
-
ing the SourceMeter to talk, a value is sent to the computer. This 
value indicates which bits in the appropriate register are set.
Program event enable registers
ENABle <NDN> or <NRf>
:STATus:MEASurement:ENABle <NDN> or <NRf> Program Measurement Event Enable  
Register
:STATus:QUEStionable:ENABle <NDN> or <NRf> Program Questionable Event Enable  
Register
:STATus:OPERation:ENABle <NDN> or <NRf> Program Operation Event Enable 
Register 
Description These commands are used to set the contents of the event enable  
registers.
Read condition registers
CONDition?
:STATus:MEASurement:CONDition? Read Measurement Condition
:STATus:QUEStionable:CONDition? Read Questionable Register
:STATus:OPERation:CONDition? Read Operation Condition
Description These query commands are used to read the contents of the condi-
tion registers.