18-88 SCPI Command Reference 2400 Series SourceMeter
®
 User’s Manual 
2420 
<n> =  -6.3 to 6.3 Set I-Source level (amps) 
-128 to 128 Set V-Source level (volts) 
DEFault 0A or 0V 
MINimum -6.3A or -128V 
MAXimum +6.3A or +128V 
2425 and 2430 DC Mode 
<n> =  -6.3 to 6.3 Set I-Source level (amps) 
-210 to 210 Set V-Source level (volts) 
DEFault 0A or 0V 
MINimum -6.3A or -210V 
MAXimum +6.3A or +210V 
2430 Pulse Mode 
<n> =  -21 to 21 Set I-Source level (amps) 
-210 to 210 Set V-Source level (volts) 
DEFault 0A or 0V 
MINimum -21A or -210V 
MAXimum +21A or +210V 
2440 
<n> =  -10.5 to 10.5 Set I-Source level (amps) 
-42 to 42 Set V-Source level (volts) 
DEFault 0A or 0V 
MINimum -10.5A or -84V 
MAXimum +10.5A or +84V
Query :STEP Query step size for sweep
:STEP? DEFault  Query *RST default level
:STEP? MINimum  Query lowest allowable level
:STEP? MAXimum  Query highest allowable level
Description This command is used to specify a step size for a linear sweep. 
When the sweep is started, the source level changes in equal steps 
from the start level to the stop level. A measurement is performed at 
each source step (including the start and stop levels).
NOTE This command cannot be used for a logarithmic sweep. Use the :POINTs 
command to set the source-measure points for a log sweep.
To avoid a setting conflicts error, make sure the step size is greater 
than the start value and less than the stop value.
The number of source-measure points in a linear sweep can be cal-
culated as follows:
Points = [(Stop - Start) / Step] + 1| 
Points = (Span / Step) + 1
An alternate way to set the source-measure points in a linear sweep 
is to simply specify the number of source-measure points in the 
sweep using the :POINts command.