User Manual
2400 Series SourceMeter
®
 User’s Manual SCPI Command Reference 18-87
Query :CENTer? Query center point for sweep
:CENTer? DEFault Query *RST default level
:CENTer? MINimum Query lowest allowable level
:CENTER? MAXimum Query highest allowable level
:SPAN? Query span for sweep 
:SPAN? DEFault Query *RST default level
:SPAN? MINimum Query lowest allowable level
:SPAN? MAXimum Query highest allowable level
Description A sweep can be configured by specifying center and span parame-
ters. By specifying a center point, you can sweep through the operat-
ing point of a device. The span determines the sweep width with the 
operating point at the center of the sweep.
For example, assume you are testing a device that operates at 10V, 
and you want to sweep from 8 to 12 volts. To do this, you would 
specify the center to be 10V and the span to be 4 volts (12 - 8).
Use the :STEP or :POINts command to specify the number of 
source-measure points in the sweep.
:CENTer and :SPAN are coupled to :STARt and :STOP. Thus, 
when center and span values are changed, the values for start and 
stop are affected as follows:
Start = Center - (Span / 2) 
Stop = Center + (Span / 2)
STEP <n>
:SOURce[1]:CURRent:STEP <n>  Specify step size (current sweep)
:SOURce[1]:VOLTage:STEP <n> Specify step size (voltage sweep)
Parameters 2400/2400-LV/2401 
<n> =  -2.1 to 2.1 Set I-Source level (amps) 
-420 to 420 Set V-Source level (volts) 
(-42 to 42, 2400-LV and 2401) 
DEFault 0A or 0V 
MINimum -2.1A or -420V (-42V, 2400-LV
 and 2401)
MAXimum +2.1A or +420V (+42V, 2400-LV
 and 2401)
2410 
<n> =  -2.1 to 2.1 Set I-Source level (amps) 
-2200 to 2200 Set V-Source level (volts) 
DEFault 0A or 0V  
MINimum -2.1A or -2200V 
MAXimum +2.1A or +2200V