18-84 SCPI Command Reference 2400 Series SourceMeter
®
 User’s Manual 
on a linear scale. With LOGarithmic selected, the source-measure 
points will be performed on a logarithmic scale.
STARt <n> 
STOP <n>
:SOURce[1]:CURRent:STARt <n> Specify start current level (current sweep)
:SOURce[1]:VOLTage:STARt <n> Specify start voltage level (voltage sweep)
:SOURce[1]:CURRent:STOP <n> Specify stop current level (current sweep)
:SOURce[1]:VOLTage:STOP <n> Specify stop voltage level (voltage sweep)
Parameters 2400/2400-LV/2401 
<n> =  -1.05 to 1.05 Set I-Source level (amps) 
-210 to 210 Set V-Source level (volts) 
(-21 to 21, 2400-LV and 2401) 
DEFault 0A or 0V 
MINimum -1.05A or -210V (-21V, 2400-LV
 and 2401) 
MAXimum +1.05A or +210V (+21V, 2400-LV
 and 2401) 
2410 
<n> =  -1.05 to 1.05 Set I-Source level (amps) 
-1100 to 1100 Set V-Source level (volts) 
DEFault 0A or 0V  
MINimum -1.05A or -1100V 
MAXimum +1.05A or +1100V 
2420 
<n> =  -3.15 to 3.15 Set I-Source level (amps) 
-63 to 63 Set V-Source level (volts) 
DEFault 0A or 0V 
MINimum -3.15A or -63V 
MAXimum +3.15A or +63V 
2425 and 2430 DC Mode 
<n> =  -3.15 to 3.15 Set I-Source level (amps) 
-105 to 105 Set V-Source level (volts) 
DEFault 0A or 0V 
MINimum -3.15A or -105V 
MAXimum +3.15A or +105V
2430 Pulse Mode 
<n> =  -10.5 to 10.5 Set I-Source level (amps) 
-105 to 105 Set V-Source level (volts) 
DEFault 0A or 0V 
MINimum -10.5A or -105V 
MAXimum +10.5A or +105V 