 User’s Manual SCPI Command Reference 18-61
RESistance:MODE <name>
[:SENSe[1]]:RESistance:MODE <name> Select ohms measurement mode
Parameters <name> = MANual Manual ohms mode
AUTO Auto ohms mode
Query :MODE? Query ohms mode
Description This command is used to select the ohms measurement mode. With 
MANual ohms selected, the user must configure the source and 
measure aspects of the operation. When the ohms function is 
selected, the ohms reading is simply the result of the V/I calculation.
Range changes cannot be made in manual ohms.
With AUTO ohms selected, the SourceMeter will be configured to 
Source I Measure V when the ohms function is selected. The current 
source value and voltage measurement range used depends on the 
ohms measurement range that is selected.
See Section 4, “Ohms Measurements,” for details on manual and 
auto ohms.
RESistance:OCOMpensated <b>
[:SENSe[1]]:RESistance:OCOMpensated <b> Control offset-compensated ohms
Parameters <b> = 1 or ON Enable offset compensation 
0 or OFF Disable offset compensation
Query :OCOMpensated? Query state of offset compensation
Description This command is used to enable or disable offset-compensated 
ohms. When using the auto ohms measurement mode, the current 
source level is automatically set. When using the manual ohms mea
-
surement mode, you must set the source (V or I) output level.
See Section 4, “Offset-compensated ohms,” for details on making 
offset-compensated ohms measurements.
NOTE Offset-compensated ohms will disable when the :MEASure? command (for 
the resistance function) or the :CONFigure:RESistance command is sent.
NOTE For the Model 2430 Pulse Mode, offset-compensated ohms cannot be 
enabled using this command. Sending this command results in error +831.