User Manual
18-38 SCPI Command Reference 2400 Series SourceMeter
®
 User’s Manual 
PASS:SOURce2 <NRf> | NDN
:CALCulate2:LIMitx:PASS:SOURce2 <NRf> | <NDN> Set sorting mode pass pattern (x = 2, 3, 5-12)
NOTE The above command to set the pass pattern for the sorting mode is invalid 
for the Model 2401 and will cause an undefined header error.
Parameters <NRf> =  0 to 7 (3-bit)  Decimal value  
0 to 15 (4-bit)  Decimal value 
<NDN> =  0 to #b111 (3-bit) Binary value 
0 to #b1111 (4-bit) Binary value 
0 to #q7 (3-bit) Octal value 
0 to #q17 (4-bit) Octal value 
0 to #h7 (3-bit) Hexadecimal value 
0 to #hF (4-bit) Hexadecimal value
Query :SOURce2? Query programmed source value 
Description This command is used to define the 3-bit or 4-bit output pattern for 
the Digital I/O Port when a test (limit 2, 3, 5-12) for the sorting mode 
passes. Note that the output value can be specified in binary, octal, 
decimal, or hexadecimal format. Use the table provided in the 
“Description” for the :SOURce command to determine the parameter 
value for the desired decimal digital output pattern.
STATe <b>
:CALCulate2:LIMit[1]:STATe <b> Control LIMIT 1 test 
:CALCulate2:LIMitx:STATe <b> Control LIMIT x test (x = 2, 3, 5-12)
Parameters <b> = 1 or ON Enable specified limit test
0 or OFF Disable specified limit test
Query :STATe? Query state of specified limit test 
Description These commands are used to enable or disable LIMIT 1, LIMIT 2, 
LIMIT 3, and LIMIT 5 to LIMIT 12 tests. Any limit test not enabled is 
simply not performed.
When a limit test is enabled, the Digital I/O port comes under control 
of limit tests. That is, the result of the testing process updates the 
output pattern on the I/O port.