5. If you chose to use a user setup as the power-on configuration, select the 
user setup number, then press ENTER.
Factory default settings
As summarized in Table 1-2, there are two sets of factory defaults, BENCH (front 
panel) and GPIB (remote). You can restore either of these default conditions as 
follows:
1. Press the MENU key, select SAVESETUP, then press ENTER.
2. From the SAVESETUP menu, select GLOBAL, then press ENTER.
3. From the GLOBAL SETUP MENU, select RESET, then press ENTER.
4. Select BENCH or GPIB defaults as desired, then press ENTER to complete 
the process.
Table 1-2 
Factory default settings
Setting
BENCH or GPIB 
default
A/D Controls:
Auto-zero
On
Line frequency No effect
Beeper On
Contact check mode
Disabled**
Contact check threshold resistance
50ohms**
Contact check event detection
Disabled**
Contact check timeout
0**
Data Store No effect
Digital output HI, HI, HI, HI
Digits 5
Fan (2420, 2425, 2430, and 2440) Output on
FCTN Power (off)
Filter: Off
Averaging type Repeat
Count
10
GPIB address No effect
Guard Cable
Limit tests:
DigOut:
Size 4-bit
Mode: Grading
Binning control Immediate
Auto clear: Disabled
Delay 0.00001 sec