User Manual
2400 Series SourceMeter
®
 User’s Manual SCPI Command Reference 18-13
[:SENSe[1]]
:VOLTage[:DC]
:PROTection
[:LEVel]? Query voltage compliance limit.
✓
:TRIPped? In voltage compliance?: 1 (yes), 0 (no). ✓
:RSYNchronize <b> Enable or disable measure and  
compliance range synchronization.
OFF
:RSYNchronize? Query range synchronization.
:RESistance Path to configure resistance: ✓
:MODE <name> Select ohms mode (MANual or AUTO). AUTO ✓
:MODE? Query ohms mode.
:OCOMpensated <b> Enable or disable offset-compensated 
ohms.
1
OFF ✓✓
:OCOMpensated? Query state of offset-compensated ohms. ✓
:RANGe Configure measurement range: ✓
[:UPPer] 
<n>|UP|DOWN
Select range by specifying the expected 
resistance reading.
2
2.1e5 ✓✓
[:UPPer]? Query range. ✓
:AUTO <b> Enable or disable auto range.
3
ON ✓✓
:AUTO? Query auto range. ✓
ULIMit <n> Set auto range upper limit.
2
✓
ULIMit? Query auto range upper limit. ✓
LLIMit <n> Set auto range lower limit.
2
✓
LLIMit? Query auto range lower limit. ✓
:HOLDoff <b> Enable/disable current range holdoff
for source memory sweeps.
✓
:HOLDoff? Query current range holdoff state.
:DELay <NRf> Set holdoff delay (sec.): 0 to 999.9999. ✓
:DELay? Query holdoff delay.
:NPLCycles <n> Specify integration rate (in line cycles): 
0.01 to 10.
4
1.0 ✓✓
:NPLCycles? Query integration rate. ✓
1. For the Model 2430 Pulse Mode, offset-compensated ohms cannot be enabled using this command. This command is invalid (error 
+831).
2. The parameter values for this command are provided with the detailed command reference information. Detailed information on SCPI 
commands follows the last command summary table in this section.
3. For the Model 2430 Pulse Mode, auto range is invalid (error +831).
4. For the Model 2430 Pulse Mode, <n> = 0.004 to 0.100.
Table 18-6 (continued)
SENSe command summary
Command Description
Default 
parameter SCPI
Source 
memory










