 User’s Manual Limit Testing 12-21
Command* Description*
:CALCulate2:LIMit[1]:COMPliance:SOURce2  
<NRf> | <NDN>
Specify limit 1 fail bit pattern. (NRf | NDN =  
pattern).
:CALCulate2:LIMitX:LOWer:SOURce2 <NRf> | 
<NDN>
Specify lower Limit X fail bit pattern for grading 
mode; X = 2, 3, 5-12 (NRf | NDN = bit pat- 
tern).
:CALCulate2:LIMitX:UPPer:SOURce2 <NRf> | 
<NDN>
Specify upper Limit X fail bit pattern for grading 
mode; X = 2, 3, 5-12 (NRf = bit pattern).
:CALCulate2:LIMitX:PASS:SOURce2 <NRf> | 
<NDN>
Specify pass pattern for sorting mode (NRf | 
NDN = bit pattern). 
:CALCulate2:LIMit[1]:STATe <state> Enable/disable Limit 1 test (state = ON or 
OFF).
:CALCulate2:LIMitX:STATe <state> Enable/disable Limit X test; X = 2, 3, 5-12 
(state = ON or OFF).
:CALCulate2:LIMit[1]:FAIL? Query Limit 1 test result (0 = pass, 1 = fail).
:CALCulate2:LIMitX:FAIL? Query Limit X test result; X = 2, 3, 5-12 (0 = 
pass, 1 = fail).
:CALCulate2:CLIMits:PASS:SOURce2 <NRf> | 
<NDN>
Specify pass bit pattern (NRf | NDN = pattern). 
Sorting mode only if limits 2, 3 and 5-12 dis- 
abled.
:CALCulate2:CLIMits:FAIL:SOURce2 <NRf> | 
<NDN>
Specify fail bit pattern for sorting mode (NRf | 
NDN = pattern).
:CALCulate2:CLIMits:PASS:SMLocation <location> Specify pass source memory location. Location 
=
NRf (memory #) or NEXT (next location).
:CALCulate2:CLIMitis:FAIL:SMLocation <location> Specify fail source memory location. (Location 
=  NRf (memory#) or NEXT (next location).
:CALCulate2:CLIMits:BCONtrol <name> Control I/O port pass/fail update. Name =  
IMMediate (at failure) or END (end of sweep).
:CALCulate2:CLIMits:CLEar  Clear test results, reset I/O port.
:CALCulate2:CLIMits:CLEar:AUTO <state> Enable/disable auto-clear (state = ON or OFF).
:CALCulate2:CLIMits:MODE <name> Select Digital I/O control mode (name =  
GRADing or SORTing).
*  LIMitX = LIMit2, LIMit3, LIMit5 through LIMit12. LIMit4 (Contact check option) not shown (see Table 18-1 and 
Appendix F).
Table 12-1 (cont.)
Limit commands