 User’s Manual Limit Testing 12-7
the Binning Control decision block. (Note that the pass condition can also be 
determined with the :CALC2:LIM<n>FAIL? query via remote.)
Immediate binning — For immediate binning, the testing process stops. The 
SourceMeter outputs the pass pattern via the Digital I/O port to the component 
handler to perform the binning operation.
End binning — For end binning, operation drops down to Another Test Cycle? 
decision block. If programmed to perform additional tests (i.e., sweep) on the DUT 
package, operation loops back up to perform the next source-measure action. 
After all programmed test cycles are successfully completed, the SourceMeter 
outputs the pass pattern to the component handler to perform the binning opera-
tion.
If configured to test another DUT package, the operation loops back to the top of 
the flowchart and waits for the start-of-test (SOT) pulse from the component han-
dler.
Fail condition
When a failure occurs, the FAIL message is displayed (and also can be read via 
remote with :CALC2:LIM<n>FAIL?), and operation proceeds to the Binning Con-
trol decision block.
Immediate binning — For immediate binning, the testing process is terminated 
when a failure occurs, and the fail pattern for that particular failure is sent to the 
component handler to perform the binning operation. Triggering is maintained for 
other instruments or components in the test system.
End binning — For end binning, the fail pattern for the first failure is stored in 
memory and operation proceeds to Another Test Cycle? decision block. If pro-
grammed to perform additional tests (i.e., sweep) on the DUT package, operation 
loops back up to perform the next source-measure action.
After all programmed test cycles are completed, the SourceMeter outputs the fail 
pattern stored in memory on the Digital I/O port. This reflects the first failure that 
occurred in the testing process for the device package. The component handler 
places the DUT in the appropriate bin.
If configured to test another DUT package, operation loops back to the top of the 
flowchart and waits for the start-of-test (SOT) pulse from the component handler.
Sorting mode
Sorting mode limits operation is detailed by the flowchart in Figure 12-5. A test is 
only performed if it is enabled. If disabled, operation proceeds to the next test. The 
following assumes the digital output of the SourceMeter is connected to a compo-
nent handler for DUT binning. (See “Binning systems,” page 12-10.)