User Manual
11-16 Triggering 2400 Series SourceMeter
®
 User’s Manual 
Figure 11-4
Measure action
Counters
Programmable counters are used to repeat operations within the trigger model 
layers. For example, if performing a 10-point sweep, the trigger counter would be 
set to 10  
(TRIGger:COUNt 10). Operation will stay in the Trigger Layer until the 10 
source-delay-measure points of the sweep are performed.
If you wanted to repeat the sweep three times, the arm counter would be set to 
three (ARM:COUNt 3). Three 10-point sweeps can then be performed for a total of 
30 source-delay-measure actions.
The maximum buffer size for the SourceMeter is 2500 readings. The product of the 
finite values of the two counters cannot exceed 2500. For example, if you set an 
arm count of two, the maximum trigger count will be 1250 (2500 / 2 = 1250). How-
ever, you can set the arm count to infinite (INF). With an infinite arm count, the max-
imum trigger count is 2500.
NOTE With front panel operation only, when a sweep is configured, the trigger 
model settings will not change until the sweep is started. After the sweep 
is finished, the trigger model will reset to the previous settings.
CONV
CONV
CONV
CONV = Reading Conversion
MEASURE
Action
Filter Process
Offsest Comp. Ohms
Autorange
CALC1, CALC2
(Repeat)