User Manual
10-18 Sweep Operation 2400 Series SourceMeter
®
 User’s Manual 
Step 2: Configure sweep
Configure the sweep as follows:
1. Press CONFIG then SWEEP.
2. Select TYPE, then press ENTER.
3. Select SRC MEMORY, then press ENTER.
4. Use the menu selections to enter the desired START memory location and 
# POINTS for the source memory sweep.
5. From the CONFIGURE SWEEPS menu, select SWEEP COUNT, press 
ENTER, then choose FINITE or INFINITE as desired.
6. Again from the CONFIGURE SWEEPS menu, choose SOURCE 
RANGING, press ENTER, then select BEST FIXED, AUTO RANGE, or 
FIXED as appropriate.
7. Press EXIT to return to normal display.
Step 3: Turn output on.
Press the ON/OFF OUTPUT key to turn the output on (OUTPUT indicator turns 
on).
Step 4: Run sweep.
To run the sweep, press the SWEEP key. After the sweep is completed, turn the 
output off by pressing the ON/OFF OUTPUT key.
Step 5: Read buffer.
Use the RECALL key to access the source-measure readings stored in the buffer. 
Use the TOGGLE to display statistical information.