User Manual
7-8 Range, Digits, Speed, and Filters 2400 Series SourceMeter
®
 User’s Manual 
NOTE After setting speed, display resolution can be changed using the DIGITS 
key.
PULSE SPEED (NPLC) – Model 2430 Pulse Mode
Press SPEED or CONFIG SPEED to display the speed choices:
0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 0.09 0.10
NOTE Display resolution for the Model 2430 Pulse Mode is not affected by the 
speed setting.
Remote speed programming
Speed commands
Table 7-3 summarizes commands to control speed. See Section 18, “Set mea-
surement speed,” for more information.
NOTE The speed setting is global, which means that setting the speed for one 
function affects the speed for the other functions. However, the speed 
setting is stored in source memory, and you can use this feature to 
change speed during a sweep. See 
Section 10 for details on using 
source memory.
Table 7-3
Speed commands
Command Description
:SENSe:CURRent:NPLCycles <n>
:SENSe:VOLTage:NPLCycles <n>
:SENSe:RESistance:NPLCycles <n>
Set amps speed (n = PLC, 0.01 to 10)*.
Set volts speed (n = PLC, 0.01 to 10)*.
Set ohms speed (n =PLC, 0.01 to 10)*.
* For the Model 2430 Pulse Mode, n = PLC, 0.01 to 0.10.