6-36 Source-Measure Concepts 2400 Series SourceMeter
®
 User’s Manual
Buffer considerations
When the SourceMeter is in the process of storing readings, configuration 
changes affect what gets stored in the buffer. These storage considerations and 
restrictions are summarized in Table 6-3.
The first column of Table 6-3 shows the SourceMeter configuration when the stor-
age process is started. The next three columns show what happens when config-
uration changes are made while the SourceMeter is storing readings.
Changing V, I, or Ω measurement function
• If you started with only a basic measurement function selected, the buffer 
will track a basic measurement function change. For example, if you started 
in volts and changed to current, the buffer will store the current readings.
• If you started with MATH, REL, and/or LIMITS enabled, the buffer will stop 
storing readings if you change the basic measurement function. Storage 
will continue if you return to the original configuration.
See Section 3, “Basic source-measure procedure,” for more information on 
selecting the measurement function.
Table 6-3
Buffer considerations
Configuration at start of 
the storage process 
What happens if 
the basic measure-
ment function (V, I, 
or Ω) is changed?
What happens if  
the MATH function  
is changed?
What happens if REL or 
LIMITS is changed?
Measure V, I, or Ω
MATH (FCTN) enabled
REL/LIMITS enabled
Buffer tracks
Buffer pauses
Buffer pauses 
V, I, or Ω stored  
MATH not stored
OK
Buffer pauses
V, I, or Ω stored  
REL or Limits not stored
MATH stored  
REL or Limits not stored
OK 