User Manual
Data flow
Data flow for front panel operation is summarized by the block diagrams provided 
in Figure 6-20. Note that if REL is enabled, the result of the rel operation is sent to 
the other blocks.
NOTE See Appendix C for remote operation data flow information.
With Math (FCTN) and Limit Tests (LIMITS) disabled (see Figure 6-20A), the 
SourceMeter displays the measurement readings. If the data store is used, these 
readings are also stored in the buffer for later recall. Statistical data for these 
readings is also available upon recall.
Figure 6-20B shows data flow when Math or Limit Tests is enabled. If Math is 
enabled, the result of the math operation is displayed. If Limit Tests is enabled, the 
raw reading along with the results of the tests (pass or fail) is displayed. As in the 
previous case, these readings can also be stored in the data store.
Figure 6-20C shows data flow when both Math and Limit Tests are enabled. The 
Math operation is performed first, and then limit tests are performed on that math 
result. The result of the math operation and the result of the limit tests (pass or 
fail) are displayed. As shown, these readings can also be stored in the data store.