User Manual
6-32 Source-Measure Concepts 2400 Series SourceMeter
®
 User’s Manual
Guard sense
When the GUARD-to-LO resistance path is less than 1kΩ, remote guard sensing 
should be used to compensate for IR drop in the GUARD test lead and/or switch 
contacts on a switching card.
Figure 6-18 was modified to create Figure 6-19A by changing the value of R3 to 
100Ω and showing the 1Ω resistance (R
TL
) of the GUARD test lead. Since the 
resistance path from GUARD-to-LO is less than 1kΩ, the IR drop in the guard test 
lead (R
TL
) becomes significant. The guard voltage applied to the bottom of R2 is 
now significantly lower than the In/Out HI voltage of the SourceMeter. As a result, 
leakage current (I
L
) flows through R2, adversely affecting the resistance measure-
ment of R1.
The guard test lead IR drop is compensated for by connecting GUARD SENSE as 
shown in Figure 6-19B. Sensing allows the guard voltage to be sensed (mea-
sured) at the resistor network for better guard voltage regulation. If the remotely 
sensed guard voltage is less than the output voltage of the SourceMeter, the 
guard voltage will be increased until the sensed guard voltage equals the output 
HI voltage.
Note that in order to ensure that guard current (I
G
) in Figure 6-19 does not exceed 
50mA, output voltage from the SourceMeter must not exceed 5V (50mA ×  100Ω = 
5V).
NOTE Guard sense operation is automatic. There is no menu selection to 
enable or disable guard sense.
For 6-wire ohms guard measurements, use the GUARD output-off state. 
For details on the GUARD output-off state, see 
Section 13, “Output-off 
states.”










