F-2 Contact Check Function 2400 Series SourceMeter
®
 User’s Manual 
Introduction
This appendix contains information on the contact check function available with 
the Models 2400-C, 2410-C, 2420-C, 2425-C, 2430-C, and 2440-C SourceMe-
ters.
Overview
Description
The contact check function prevents measurements that may be in error due to 
excessive resistance in the force or sense leads when making remotely sensed 
(Kelvin) measurements. There are many potential sources for this resistance, 
including poor contact at the DUT, failing relay contacts on a switching card, and 
wires that are too long or thin. This is shown schematically in Figure F-1, where 
R
C
is the resistance of the mechanical contact at the DUT, and R
S
is the series 
resistance of relays and cables. The contact check process performs a differential 
resistance measurement between each source lead and it's corresponding sense 
(Kelvin) lead. The resultant resistance is then compared to the programmed limit 
and a pass or fail result is determined. This process takes place in as short as 
350uS and delivers no energy to the device under test. 