User manual
A-8 Specifications 2400 Series SourceMeter
®
 User’s Manual 
Cases V and VI: Measure one function, Output Auto-Off disabled, and 
no source setting changes
Figure A-5
Case V Timing Diagram
Auto-Zero: Enabled 
Trigger Latency:  500µsec max
Figure A-6
Case VI Timing Diagram
Auto Zero: Disabled 
Trigger Latency: 100µsec max
The source turns on as soon as the output is turned on and remains on until the 
source is turned off. As shown in the two timing diagrams, the static source 
remains on for every measurement cycle. The Source-Delay portion of the SDM 
cycle is omitted. With Trigger Delay set to zero, Trigger Latency is the time from 
when the trigger event occurs to when the SourceMeter begins an A/D conver-
sion.
Trigger
Event
Source On
Trigger
Delay
Trigger
Latency
A/D
Conversion
(ref zero phase)
A/D
Conversion
(ref phase)
A/D
Conversion
(signal phase)
Trigger
Event
Source On
Trigger
Delay
Trigger
Latency
A/D
Conversion
(signal phase)