User manual
2400 Series SourceMeter
®
 User’s Manual Specifications A-5
Source On Time ≅  Source Configuration + Source Delay + (3 × A/D Conversion) 
+ Firmware Overhead
Example:  Source Delay = 0µsec 
NPLC Setting = 0.01 PLC 
Power Line Frequency = 60Hz
Source On Time≅ 50µsec + 0 + [(3× 0.01 × 1/60) + 185µsec] + 1.6msec 
≅  2.9msec for Source V 
≅  3.25msec for Source I
Case II: Auto Zero enabled and measuring two functions
Figure A-2
Case II Timing Diagram
Trigger Latency:  225µsec max 
Source Configuration:  50µsec max 
A/D Conversion: [NPLC Setting × (1 / power line frequency)] + 185µsec 
Firmware Overhead:  2.3msec for Source V 
2.65msec for Source I
Source On Time ≅  Source Configuration + Source Delay + (4 × A/D Conversion) 
+ Firmware Overhead
Example:  Source Delay = 0µsec 
NPLC Setting = 0.06 PLC 
Power Line Frequency = 60Hz
Source On Time≅  50µsec + 0 + [(4 × 0.06 × 1/60) + 185µsec] + 2.6msec 
≅  7.1msec for Source V 
≅  7.45msec for Source I