User manual
2400 Series SourceMeter
®
 User’s Manual SCPI Command Reference 18-105
Description This command is used to enable or disable remote sensing. When 
using remote sensing, 4-wire connections to the DUT are required.
Sourcing voltage — With remote sensing enabled, the output volt-
age is sensed (measured) at the DUT. If the sensed voltage is lower 
than the programmed amplitude, then the V-Source increases the 
voltage until the sensed voltage is the same as the programmed 
amplitude. This compensates for IR drop in the OUTPUT test leads.
With remote sensing disabled, the output voltage is sensed at the 
output connectors.
Measuring voltage — With remote sensing enabled, voltage mea-
surements are made at DUT. This eliminates any voltage drops that 
may be in the test leads between the SourceMeter and the DUT.
With remote sensing disabled, voltage measurements are performed 
at the output connectors of the instrument.
Measuring resistance — With remote sensing enabled, 4-wire 
resistance measurements can be made.
NOTE See Section 2 for details on sensing.
Select guard mode
GUARd <name>
:SYSTem:GUARd <name> Select guard mode
Parameters <name> = OHMS Ohms guard mode
CABLe  Cable guard mode
Query :GUARd? Query guard mode
Description This command is used to select the guard mode. OHMS guard is a 
low-impedance guard drive used for in-circuit resistance 
measurements. Ohms guard is not available for SourceMeters that 
have 1A, 3A and/or 10A ranges (source or measure). CABLE guard 
provides a high-impedance guard drive that is used to eliminate 
leakage currents in cabling and test fixtures.
When performing 6-wire ohms guard measurements, use the 
GUARD out-put state. The OUTPut [1]:SMODe GUARd command is 
used to select the GUARD output-off state.
NOTE See Section 2 for details on guarding.