2400 Series SourceMeter
®
 User’s Manual SCPI Command Reference 18-103
DISable <list>
:STATus:QUEue:DISable <list> Disable messages for Error Queue
Parameters <list> = (numlist)
where numlist is a specified list of messages that you wish to disable 
for the Error Queue.
Query :DISable? Query list of disabled messages
Description On power-up, all error messages are enabled and will go into the 
Error Queue as they occur. Status messages are not enabled and 
will not go into the queue. This command is used to specify which 
messages you want disabled. Disabled messages are prevented 
from going into the Error Queue.
SYSTem subsystem
The SYSTem subsystem contains miscellaneous commands that are summarized in 
Table 18-9, SYSTem command summary.
Default conditions
PRESet
:SYSTem:PRESet Return to :SYSTem:PRESet defaults
Description This command returns the instrument to states optimized for front 
panel operation. :SYSTem:PRESet defaults are listed in the SCPI 
tables (
Table 18-1 through Table 18-11).
POSetup
:SYSTem:POSetup <name> Program power-on defaults
Parameters <name> = RST Power-up to *RST defaults
PRESet Power-up to :SYSTem:PRESet 
defaults
SAV0 Power-up to setup stored at memory 
location 0
SAV1 Power-up to setup stored at memory 
location 1
SAV2 Power-up to setup stored at memory 
location 2
SAV3 Power-up to setup stored at memory 
location 3
SAV4 Power-up to setup stored at memory 
location 4
Query :POSetup? Query power-on setup 