User manual
2400 Series SourceMeter
®
 User’s Manual SCPI Command Reference 18-83
Configure voltage and current sweeps
There are two methods to configure the start and stop levels of a sweep. You can use 
either the :STARt and :STOP commands or you can use the :CENTer and :SPAN 
commands.
NOTE In order to run a sweep, the selected source must be in the sweep sourcing 
mode and the trigger count should be the same as the number of source-
measure points in the sweep.
Use the :FUNCtion:MODE command to select the current or voltage source function. 
(See “Select function modes,” page 18-72.) Use the :CURRent:MODE or VOLT-
age:MODE command to select the SWEep sourcing mode. (See “Select sourcing 
mode,” page 18-73.) The trigger count is set using the TRIGger:COUNt command. 
(See “TRIGger subsystem,” page 18-116.)
RANGing <name>
:SOURce[1]:SWEep:RANGing <name> Select source ranging mode
Parameters <name> = BEST Use the best fixed mode
AUTO Use the most sensitive source 
range for each sweep level
FIXed Use the present source range for 
the entire sweep
Query :RANGing? Query source ranging mode
Description This command is used to select the source ranging mode for 
sweeps. With BEST selected, the SourceMeter will select a single 
fixed source range that will accommodate all the source levels in the 
sweep. For front panel operation, this is the BEST FIXED option.
With AUTO selected, the SourceMeter will automatically go to the 
most sensitive source range for each source level in the sweep. For 
front panel operation, this is the AUTO RANGE option.
With FIXed selected, the source remains on the range that it is pres-
ently on when the sweep is started. For sweep points that exceed the 
source range capability, the source will output the maximum level for 
that range. For front panel operation, this is the FIXED option.
SPACing <name>
:SOURce[1]:SWEep:SPACing <name> Select scale for sweep 
Parameters <name> = LINear Linear scale
LOGarithmic Logarithmic scale
Query :SPACing? Query scale for sweep
Description This command is used to select the scale for the sweep. With LINear 
selected, the source-measure points in the sweep will be performed 