User manual
16-4 Common Commands 2400 Series SourceMeter
®
 User’s Manual 
Use the *SAV command to save the present instrument setup configuration in 
memory for later recall. Any control affected by *RST can be saved by the *SAV 
command. The *RCL command is used to restore the instrument to the saved 
setup configuration. Five setup configurations can be saved and recalled.
The SourceMeter ships from the factory with SYSTem:PRESet defaults loaded 
into the available setup memory. If a recall error occurs, the setup memory 
defaults to the  
SYSTem:PRESet values.
**RST — reset
When the *RST command is sent, the 
SourceMeter performs the following operations:
• Returns the SourceMeter to the GPIB default conditions (see “Default” col-
umn of SCPI tables in Section 18).
• Cancels all pending commands.
• Cancels response to any previously received *OPC and *OPC? com-
mands.
*TRG — trigger
Use the *TRG command to issue a 
GPIB trigger to the SourceMeter. It has the same effect as a group execute trigger 
(GET).
Use the *TRG command as an event to control operation. The SourceMeter 
reacts to this trigger if BUS is the programmed arm control source. The control 
source is programmed from the TRIGger subsystem.
NOTE Details on triggering are covered in Section 11.
**TST? — self-test query
Use this query command to perform a 
checksum test on ROM. The command places the coded result (0 or 1) in the Out-
put Queue. When the SourceMeter is addressed to talk, the coded result is sent 
from the Output Queue to the computer.
A returned value of zero (0) indicates that the test passed, and a value of one (1) 
indicates that the test failed.
Return SourceMeter to GPIB defaults
Send bus trigger to SourceMeter
Run self test and read result