User manual
10-24 Sweep Operation 2400 Series SourceMeter
®
 User’s Manual 
Pulse Mode sweeps (Model 2430 only)
When performing a sweep with the Model 2430 in the Pulse Mode, each step or 
point of the sweep is made up of a pulse period. In general, a pulse period con-
sists of the pulse width (output on-time) and the output off-time. Figure 10-5 
shows an example of a 3-step linear staircase sweep for the Pulse Mode (start = 
1V, stop = 3V, step = 1V). A measured reading is performed for each of the three 
pulse steps. Readings are not performed for the output off-times.
For normal (DC Mode) sweep operation, the time duration spent on each step 
consists of the set source delay and the time it takes to perform the measurement. 
For the Pulse Mode, source delay is not used. The pulse width consists of the 
pulse width delay, internal overhead time and the pulse signal measurement. The 
output off-time typically consists of zero and reference measurements (used to 
calculate accurate pulse readings), internal overhead time, and an optional pulse 
delay.
NOTE Details on Pulse Mode operation for the Model 2430 are provided in 
Section 5.










