2400 Series SourceMeter
®
 User’s Manual Range, Digits, Speed, and Filters 7-3
Auto ranging
NOTE For the Model 2430, auto ranging cannot be performed while in the Pulse 
Mode. Therefore, all the following information on auto ranging is not 
applicable to the Model 2430 Pulse Mode.
For the Source V Measure I, Source I Measure V, and Ohms configurations, press 
AUTO RANGE to enable auto ranging. The AUTO annunciator turns on when 
auto ranging is selected. With auto ranging selected, the instrument automatically 
chooses the best range to measure the applied signal. Note that the highest avail-
able range is dependent on the corresponding compliance setting. 
Auto range considerations
If the SourceMeter has to autorange, the SDM (source-delay-measure) cycle is 
repeated to take the measurement on the new range. This means that any pro-
grammed source delay will be applied each time the unit has to autorange. For 
example, if a 1 second source delay is programmed, the unit could take 2 or more 
seconds to complete a reading if it has to change ranges.
The autorange algorithm used in the SourceMeters is as follows: If the reading 
reaches 105% of the present range, the unit will go up 3 ranges, or to the highest 
range possible if it can't go up 3 ranges. The unit will take another reading, then 
decide whether it needs to continue going up in range or pick the right range 
based on the reading.  If the reading is 10%, 1%, or 0.1% of the present range, it 
will go down by 1, 2, or 3 ranges based on the reading.
Auto range change mode
The auto range change mode determines how the instrument performs auto rang-
ing. In the SINGLE mode, the SourceMeter will auto range only after first taking a 
reading. In the MULTIPLE mode, the SourceMeter will auto range up on compli-
ance in the Delay phase of the Source-Delay-Measure (SDM) cycle, thereby mini-
mizing the possibility that a SourceMeter will be in compliance in a multiple-
SourceMeter system. The SourceMeter can downrange only once a reading has 
been taken.
NOTE See Section 6, “Source-delay-measure cycle,” for more SDM informa-
tion.
With the auto range change mode set to MULTIPLE, you can also program the 
soak time, which specifies the amount of time after the first point of a sweep that 
the unit will sit in a loop actively auto ranging up and down to allow a multiple 
SourceMeter configuration to settle. This process will occur only during the first 
SDM cycle after the initial sweep trigger. (See Section 11, “Trigger models.”) This 
feature is especially useful for situations with long DUT settling times (such as low 
current measurements) when several down-range change cycles from the higher 
ranges are required.