6-4 Source-Measure Concepts 2400 Series SourceMeter
®
 User’s Manual
Compliance examples
When the SourceMeter goes into real compliance, the Cmpl label for the compli-
ance display will flash. When the SourceMeter goes into range compliance, the 
units label (“mA”) will flash instead. For the following examples, labels in boldface 
type indicate that they are flashing.
Measurement Range:100mA
Compliance Setting: Cmpl: 075.000 mA  Flashing Cmpl indicates that real compliance 
has occurred. The output is clamped at 75mA.
Measurement Range:10µA
Compliance Setting: Cmpl: 075.000 µA  Flashing mA indicates that range compliance 
has occurred. The output is clamped at 
10.5µA.
Determining compliance limit
Table 6-2 provides examples (Model 2400) for determining the actual compliance 
limit. For the first three entries in the table, the compliance setting is 150V. On the 
200V measurement range, the actual compliance is 150V (compliance setting < 
2425/2430 2440
Measure 
range
Maximum 
compliance 
value
Measure 
range
Maximum 
compliance 
value
200mV
2V
20V
100V
±210mV
±2.1V
±21V
±105V
200mV
2V
10V
40V
±210mV
±2.1V
±10.5V
±42V
10µA
100µA
1mA
10mA
100mA
1A
3A/10A
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
 **
10µA
100µA
1mA
10mA
100mA
1A
5A
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
±5.25A
 * 2400 only
** ±3.15A (2425 and 2430 DC mode)
 ±10.5A (Pulse mode; 2430 only)
Table 6-1 (cont.)
Compliance limits