 User’s Manual Source-Measure Concepts 6-3
When in range compliance, the source output clamps at the maximum compliance 
value for the fixed measurement range (not the compliance value). For example, if 
compliance is set to 1V and the measurement range is 200mV, output voltage will 
clamp at 210mV. In this situation, the units in the compliance display field will 
flash. For example, with the following display: Vcmpl: 10mA, the “mA” units indica-
tion will flash.
Maximum compliance values
The maximum compliance values for the measurement ranges are summarized in 
Table 6-1.
Table 6-1 
Compliance limits
2400/2400-LV/2401 2410 2420
Measure 
range
Maximum 
compliance 
value
Measure 
range
Maximum 
compliance 
value
Measure 
range
Maximum 
compliance 
value
200mV
2V
20V
200V*
±210mV
±2.1V
±21V
±210V
200mV
2V
20V
1000V
±210mV
±2.1V
±21V
±1.1kV
200mV
2V
20V
60V
±210mV
±2.1V
±21V
±63V
1µA
10µA
100µA
1mA
10mA
100mA
1A
±1.05µA
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
1µA
10µA
100µA
1mA
20mA
100mA
1A
±1.05µA
±10.5µA
±105µA
±1.05mA
±21mA
±105mA
±1.05A
10µA
100µA
1mA
10mA
100mA
1A
3A
±10.5µA
±105µA
±1.05mA
±10.5mA
±105mA
±1.05A
±3.15A