 User’s Manual 
Binning failure indications (grading mode)
As previously discussed, it is possible to place a fail pattern on the Digital I/O Port 
when contact check fails. It is important to note that, depending on whether 
source auto clear is enabled and when the bin control is done, certain issues arise 
when running a test with a normal voltage or current sweep and a source memory 
sweep. Four possible effects on sweep operation may occur when using the digi-
tal output and contact check. Refer to “Binning systems,” page 12-10, for a 
description of binning and to “Front panel output configuration,” page 13-7, for 
details on source auto off functionality.
When performing a source memory sweep with the SourceMeter, contact check is 
performed before each point in that sweep only when 4-wire remote sensing is 
enabled. To save test cycle time, the contact check is performed before the mem-
ory location for the sweep point is recalled. If there is a failure in any of the four 
modes below, the sweep memory location containing the open lead will not be 
recalled from the memory.
Auto clear on, end binning
Source auto clear on: :SOUR:CLE:AUTO ON 
Bin control at end of sweep: :CALC2:CLIM:BCON END
If a point in a sweep operation fails contact check, the output is turned off and will 
remain off until the end of the sweep points. The triggers will be supplied and 
accepted to maintain timing. The memory for each of the points after the failure 
will not be recalled, and no further contact checks will be performed until the end 
of the sweep points. If the sweep is to be repeated, the new sweep will start with a 
contact check, and normal measurement operation will resume.
Auto clear on, immediate binning
Bin control at end of each point sweep: :CALC2:CLLM:BCON IMM
If contact check detects an open lead, the output is not turned on for that particu-
lar location, and that sweep point is not recalled and will be skipped. The sweep 
proceeds to the next sweep point, and a contact check test is performed.
Auto clear off, end binning
Source auto clear off: :SOUR:CLE:AUTO OFF 
Bin control occurring at end of sweep: :CALC2:CLIM:BCON END
The output is kept on throughout the entire sweep. If an open lead is found during 
the sweep, the output is turned off, no more contacts are checked, sweep memory 
locations are not recalled, and triggers are maintained until the last sweep point. If 
multiples of the same sweep are desired, the following sweep after the open lead 