User Manual
18-80 SCPI Command Reference 2400 Series SourceMeter
®
 User’s Manual 
80 Set limit to 80V 
100  Set limit to 100V  
120  Set limit to 120V  
160  Set limit to 160V  
161 to 210 Set limit to 210V (NONE) 
DEFault Set limit to 210V (NONE) 
MINimum Set limit to 20V 
MAXimum Set limit to 210V (NONE)
2401  
<n> =  -40 to 40 Specify V-Source limit 
20 Set limit to 20V 
DEFault Set limit to 40V (NONE) 
MINimum Set limit to 20V 
MAXimum Set limit to 40V (NONE)
2410  
<n> =  -1100 to 1100 Specify V-Source limit 
20 Set limit to 20V 
40 Set limit to 40V 
100 Set limit to 100V 
200 Set limit to 200V 
300 Set limit to 300V 
400 Set limit to 400V 
500 Set limit to 500V 
501 to 1100 Set limit to NONE 
DEFault Set limit to 1100V (NONE) 
MINimum Set limit to 20V 
MAXimum Set limit to 1100V (NONE) 
2420 
<n> =  -63 to 63 Specify V-Source limit 
6 Set limit to 6V 
12 Set limit to 12V 
18 Set limit to 18V 
24 Set limit to 24V 
30 Set limit to 30V 
36 Set limit to 36V 
48 Set limit to 48V 
49 to 63 Set limit to NONE 
DEFault Set limit to 63V (NONE) 
MINimum Set limit to 6V 
MAXimum Set limit to 63V (NONE) 
2425 and 2430 
<n> =  -105 to 105 Specify V-Source limit 
10 Set limit to 10V 
20 Set limit to 20V 
30 Set limit to 30V 
40 Set limit to 40V 