 User’s Manual SCPI Signal Oriented Measurement Commands 17-5
Note that if auto output-off is disabled (:SOURce1:CLEar:AUTO 
OFF), then the output must be turned on before you can perform 
a :READ?. The output will then remain on after all source-
measure operations are completed.
If auto output-off is enabled (:SOURce1:CLEar:AUTO ON), then 
the output will automatically turn on at the beginning of each SDM 
(source-delay-measure) cycle and turn off after each 
measurement.
NOTE For the Model 2430 Pulse Mode, the output will turn on when the 
:READ? command is sent.
See Appendix C, “Data Flow,” for a detailed explanation on how data 
flows through the various operation blocks of the SourceMeter. It clarifies 
the type of readings that are acquired by the various commands to read 
data.
:MEASure[:<function>]?
Parameters <function> = CURRent[:DC]Amps function
VOLTage[:DC]Volts function
RESistanceOhms function
Description This command combines other signal oriented measurement 
commands to perform a “one-shot” measurement and acquire the 
reading. Note that if a function is not specified, the measurement 
will be done on the function that is presently selected.
When this command is sent, the following commands execute in 
the order that they are presented.
• :CONFigure:<function>
•:READ?
When :CONFigure is executed, the instrument goes into a “one-
shot” measurement mode. See :CONFigure for more details.
When :READ? is executed, its operations will then be performed. 
In general, another :ABORt is performed, then an :INITiate, and 
finally a FETCh? to acquire the readings. See :READ? for more 
details.
When :MEASure? is sent, the source turns on and a single 
measurement is performed. If auto output-off is enabled 
(:SOURce1:CLEar:AUTO ON), then the output will turn off after 
the measurement is completed. If auto output-off is disabled 
(:SOURce1:CLEar:AUTO OFF), then the output will remain on 
after the measurement is completed.