User Manual
15-16 Status Structure 2400 Series SourceMeter
®
 User’s Manual 
Condition registers
As Figure 15-1 shows, each status register set (except the Standard Event Regis-
ter set) has a condition register. A condition register is a real-time, read-only regis-
ter that constantly updates to reflect the present operating conditions of the 
instrument. For example, while the SourceMeter is in the idle state, bit B10 (Idle) 
of the Operation Condition Register will be set. When the instrument is taken out 
of idle, bit B10 clears.
The commands to read the condition registers are listed in Table 15-5. For details 
on reading registers, see “Reading registers,” page 15-6.
Event registers
As Figure 15-1 shows, each status register set has an event register. When an 
event occurs, the appropriate event register bit sets to 1. The bit remains latched 
to 1 until the register is reset. Reading an event register clears the bits of that reg-
ister. *CLS resets all four event registers.
The commands to read the event registers are listed in Table 15-6. For details on 
reading registers, see “Reading registers,” page 15-6.
Table 15-5
Condition register commands
Command Description
:STATus:OPERation:CONDition?
:STATus:MEASurement:CONDition?
:STATus:QUEStionable:CONDition?
Read Operation Condition Register.
Read Measurement Condition Register.
Read Questionable Condition Register.
Table 15-6
Event register commands
Command Description Default
*ESR?
:STATus:OPERation:[:EVENt]?
:STATus:MEASurement:[:EVENt]?
:STATus:QUEStionable:[:EVENt]?
Read Standard Event Status Register.
Read Operation Event Register.
Read Measurement Event Register.
Read Questionable Event Register.
(Note)
Note: Power-up and *CLS resets all bits of all event registers to 0. STATus:PRESet has no effect.










