User Manual
11-24 Triggering 2400 Series SourceMeter
®
 User’s Manual 
Output triggers
All three output triggers are available in the trigger layer for the Pulse Mode. If 
enabled, the output triggers occur where indicated in the trigger model.
Invalid trigger settings
Front panel operation
For front panel operation, most trigger settings are done from the “CONFIGURE 
TRIGGER” menu which is accessed by pressing CONFIG and then TRIG. For 
Pulse Mode operation, the following settings for the “TRIG LAYER” are ignored.
DELAY — Trigger delay is not used. Therefore, the trigger DELAY setting is 
ignored.
TRIGGER OUT EVENTS — For the DC Mode, source delay is set from the 
source configuration menu, which is accessed by pressing CONFIG and then 
SOURCE V or I. When the Pulse Mode is selected, the source delay menu items 
(DELAY and AUTO DELAY) are replaced by the pulse DELAY and PULSE 
WIDTH menu items.
Remote operation
The following commands are not valid in the Pulse Mode and are therefore 
ignored:
:TRIGger:DELay — Trigger delay is not used in the Pulse Mode. 
:SOURce:DELay and :SOURce:DELay:AUTO — Source delay is not used in the 
Pulse Mode.