User Manual
2400 Series SourceMeter
®
 User’s Manual Triggering 11-9
> TRIG-IN TLINK LINE — Select the input line (#1, #2, #3, or #4) 
for the Trigger Link.
> EVENT DETECT BYPASS — Set the bypass for the Source 
Event Detector. With ONCE, operation will loop around the 
Source Event Detector. With NEVER selected, operation will 
wait for an input 
trigger.
> TRIGGER IN EVENTS — Enable (ON) or disable (OFF) trig-
ger-in events (SOURCE, DELAY, and MEASURE). With a trig-
ger-in event ON, operation will wait at that event for an input 
trigger. With the trigger-in event OFF, operation will not wait. It 
will simply continue and perform the appropriate action.
NOTE For the Model 2430 Pulse Mode, the DELAY and MEASURE events are 
not used. Therefore, the settings for those events are ignored.
– TRIGGER OUT — Use to configure the trigger layer output trigger:
/ LINE — Select the Trigger Link line for the output trigger; line #1, 
#2, #3, or #4.
/ EVENTS — Enable (ON) or disable (OFF) output triggers that 
occur after the source, delay, and measure actions.
NOTE For the Model 2430 Pulse Mode, an output trigger can only occur after 
the measure action. Therefore, the output trigger settings for the source 
and delay actions are ignored.
– DELAY — Specify the time delay (in seconds) for the trigger delay.
NOTE For the Model 2430 Pulse Mode, trigger delay is not used. Therefore, the 
trigger delay setting is ignored.
– COUNT — Specify the trigger count.
• HALT — Use to return the SourceMeter to the idle state. HALT does not 
turn off the output. The programmed source level will still be available at the 
OUTPUT terminals. The following actions will take the SourceMeter out of 
idle:
– Turn the output off and then on again.
– Re-select the arm or trigger event.
– Exit from the menu structure, and then re-enter it by pressing CONFIG 
and then TRIG.