User’s Manual Pulse Mode Operation (Model 2430 only) 5-3
Pulse characteristics
NOTE For the purpose of discussion, positive polarity pulses are shown in the
following illustrations. Keep in mind that the Model 2430 can output neg
-
ative pulses.
As shown in Figure 5-1, a pulse period consists of an output on-time (pulse width)
and an output off-time. Depending on how the SourceMeter is configured, the
pulse width can be as short as 150µsec or as long as 5msec. With measurement
enabled, the output off-time can be as short as 2.9msec or as long as
~10,000sec. The shortest output off time is 1.1msec with measurements disabled.
See Table 5-1, “SourceMeter pulse mode timing summary” for details.
Figure 5-2 shows the components that make up the pulse width and the output
off-time.
NOTE See the Model 2430 specifications in Appendix A.
Figure 5-2
Pulse-measure timing for default source-measure configuration.
80µs
Meas
Sig
2.9ms
Meas
Ref &
Zero
PD
Delay = Pulse width delay, used to achieve pulse width setting.
80µs = Minimum pulse width overhead
Meas Sig = Signal measurement
2.9ms = Minimum output off-time overhead
Meas Ref & Zero = Reference and zero measurement
PD = Pulse delay setting, used to determine time between pulses.
0V or 0A
Delay
Pulse Width
(Output On-Time)
Output Off-Time