User manual
4-2 Ohms Measurements 2400 Series SourceMeter
®
User’s Manual
Ohms configuration menu
NOTE For the Model 2430 Pulse Mode, offset-compensated ohms cannot be
enabled from the ohms configuration menu. However, offset-compen
-
sated ohms is available as a math function (Section 8, “Math opera-
tions”).
To access the ohms configuration menu, press CONFIG then OHMS. Use
Section 1, “Rules to navigate menus,” to select the various items in the menu tree,
which is shown in Figure 4-1. Menu items include:
• SOURCE — Select AUTO or MANUAL source mode.
• SENSE MODE — Select 2-WIRE or 4-WIRE sense mode.
• GUARD — Choose OHMS or CABLE guard.
• OFFSET COMPENSATION — Enable or disable offset-compensated
ohms (not valid for Model 2430 Pulse Mode).
• SRC RDBK — Enable or disable source readback mode.
The following paragraphs discuss each of these aspects in detail.
Figure 4-1
Ohms configuration menu tree
CONFIG
Ω
GUARD
SOURCE
SENSE
MODE
OFFSET
COMPENSATION
SRC
RDBK
MANUALAUTO 4-WIRE2-WIRE CABLEOHMS DISABLEENABLE DISABLEENABLE